The instrument is designed to verify the current transformer for instruments below grade 0.02, and the secondary current is 1 ~ 60mA. It is specially designed for the error detection of micro instrument transformers. In the design and production, the anti-interference ability and the operation simplicity have been repeatedly demonstrated, with better technical indicators and stability performance
Overview
(2) Dynamic synchronous data acquisition technology: it can quickly track the voltage regulator and display the data stably;
(3) Easy to operate;
(4) Ultra-wide working range of current transformer measuring 1%~600%;
(5) With RS232 interface it can be operated online by computer.
Models
Standard Current Transformer | Accuracy |
0.005level
|
Primary current (A) | 100、75、30、25、20、15、10、6、5、3、2.5、1.5、1、0.6、0.3(6 times overload the maximum current of the range above20A is 120A); | |
Secondary current (mA) | 1、1.5、2、2.5、3、10/3、4、5、6、20/3、7.5、10、15、50/3、20、25、30、40、50、60; | |
Transformer Calibrator: (error display) |
Indication accuracy |
2%
|
Error resolution |
0.001%
|
|
Power supply |
10V ,50Hz
|
|
Current & Voltage Generator | Maximum output power |
约300VA
|
With fine adjustment the adjustment fineness is better than 0.1% | ||
Load Resistance (Ω) | 0、5、10、15、20、25、30、40、50、60、70、75、80、100、150、200、250、500、1000、1500、2000 |