The instrument is designed to verify the current transformer for instruments below grade 0.02, and the secondary current is 1 ~ 60mA. It is specially designed for the error detection of micro instrument transformers. In the design and production, the anti-interference ability and the operation simplicity have been repeatedly demonstrated, with better technical indicators and stability performance

Overview

Characteristics and Functions
(1) User self-calibration function: the accuracy of the calibrator can be easily adjusted to the technical indicators when leaving the factory through the operation of the panel buttons;
(2) Dynamic synchronous data acquisition technology: it can quickly track the voltage regulator and display the data stably;
(3) Easy to operate;
(4) Ultra-wide working range of current transformer measuring 1%~600%;
(5) With RS232 interface it can be operated online by computer.

Models

Standard Current Transformer Accuracy
0.005level
Primary current (A) 100、75、30、25、20、15、10、6、5、3、2.5、1.5、1、0.6、0.3(6 times overload the maximum current of the range above20A is 120A); 
Secondary current (mA) 1、1.5、2、2.5、3、10/3、4、5、6、20/3、7.5、10、15、50/3、20、25、30、40、50、60; 

Transformer Calibrator: (error display)

Indication accuracy
2%
Error resolution
0.001%
Power supply
10V ,50Hz
Current & Voltage Generator Maximum output power
约300VA
With fine adjustment the adjustment fineness is better than 0.1%
Load Resistance (Ω) 0、5、10、15、20、25、30、40、50、60、70、75、80、100、150、200、250、500、1000、1500、2000